In this work we present the new target assembly developed for solid samples irradiation by protons with different energies at maximum of 11 MeV and 40 μA for current. The new target was designed and manufactured for conventional target changer (carousel) of 11 MeV Eclipse RD medical cyclotron. The MCNP simulation of the beam profile after degradation was performed and testing irradiation of the target assembly was successfully carried out showing the target suitability for such type of cyclotron facilities.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.