Capacitance measurements at the National Institute of Standards (NIS), Egypt, are traceable to the Bureau International des Poids et Mesures (BIPM). It calibrates the main NIS standard capacitors, AH11A. In this paper, traceability of the BIPM capacitance measurements could be used to evaluate a new accurate measurement method through an Ultra-Precision Capacitance Bridge. The new method is carefully described by introducing some necessary equations and a demonstrating chart. Verification of this new method has been realized by comparing its results for the 10 pF and 100 pF capacitance standards with the results obtained by the conventional substitution method at 1 kHz and 1.592 kHz. The relative differences between the two methods are about 0.3 µF/F, which reflect the accuracy of the new measurement method. For higher capacitance ranges, the new measurement method has been applied for the capacitance measurements up to 1 μF at 1 kHz. The relative differences between the two methods are in the range of 5.5 µF/F on the average which proves the acceptable accuracy and the reliability of the new method to be used.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.