Growing interest in the time-of-flight positron emission tomography (TOF PET) prompts the study of a new Photonis XP20D0 photomultiplier, equipped with a screening grid at the anode, in application to a fast timing with LSO and LaBr 3 crystals. The high time resolution of 200±4 and 210±4 ps was obtained for 511 keV annihilation quanta using LaBr 3 and LSO crystals in the coincidence experiment with a small BaF 2 crystal, respectively. It reflects an importance of the grid and high quantum efficiency of the XP20D0. A high-time resolution observed in the present experiments makes good prospects for a development of TOF PET.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.