Abstract

In this research, new graphite paste electrodes modified with silicon dioxide (SiO2) nano particle are constructed and used for the determination of Viagra drug (Sildenafil citrate, SILC). The electrodes are constructed by preparing ion-pair for (SILC) with phosphotungstic acid (PTA) or phosphomolybdic (PMA) acid using dibutyl phthalate (DBP) as a plasticizer. These electrodes show good sensitivity towards SILC with linear range of (1.0´10-7-1.0´10-2) M, correlation coefficient, (0.9990) and the life time (120) days for both electrodes, limit of detection (5.177´10-8 and 5.026´10-8) M, optimum temperature range (23-50 and 23-65)oC , slope (57.33 and 58.01) mV/decade and optimum pH range (2-6 and 2-5), for SILC-PTA and SILC-PMA electrodes respectively. These electrodes were successfully applied for determination of SILC in pure and pharmaceutical preparation form (tablets) with recovery of not less than 98% .
 
 http://dx.doi.org/10.25130/tjps.24.2019.109

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.