Abstract

We present a new method for torque magnetometry by using a commercially available membrane-type surface-stress sensor (MSS). This sensor has a silicon membrane supported by four beams in which piezoresistive paths are integrated. Although originally developed as a gas sensor, it can be used for torque measurement by modifying its on-chip wiring. We demonstrate the magnetic-torque measurement of submillimeter-sized crystals at a low temperature and in strong magnetic fields. This MSS can observe de-Haas-van-Alphen oscillation, which confirms that it can be an alternative tool for self-sensitive microcantilevers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.