Abstract

First derivative spectrophotometric methods have been developed for the assay of Bumetanide using Shimadzu UV-1800 Model double beam UV-VIS spectrophotometer in reagents such as hydrochloric acid, phosphate buffer pH 3.6, phosphate buffer pH 2.0 and distilled water. Bumetanide has shown linearity 10-70 µg/ml in all the methods and the linear regression equations are found to be y = 0.0015x - 0.0013 (R² = 0.9993), y= 0.0007x - 0.0004 (R² = 0.9991), y = 0.0014x - 0.0015 (R² = 0.9993) and y = 0.0014x - 0.0015 (R² = 0.9995) in hydrochloric acid, phosphate buffer pH 3.6, phosphate buffer pH 2.0, and distilled water respectively. The methods are validated as per ICH guidelines and found to be precise, simple and accurate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.