Abstract
We designed a depth-encoding positron emission tomography (PET) detector by using a bottom and side readout method with a cross-arranged scintillator array. To evaluate the characteristics of the novel detector module, we used the DETECT2000 simulation tool to perform the optical photon transport in the crystal array. The detector module consists of an M (column) × N (row) cross-arranged crystal array composed of M/3 sub-arrays consisting of N × 3 crystals. The second column of the sub-array is arranged perpendicular to the first and the third columns. The crystal is optically coupled to the crystals of the other columns; however, the surfaces between the crystals in the same column are treated as reflectors. A 6 × 5 crystal array consisting of two sub-arrays was considered for proof of concept. The two multi-pixel photon counter (MPPC) arrays are coupled to the bottom and one side of the crystal array, respectively. The x-y position is determined by the bottom MPPC array, and the side MPPC array gives depth information. All pixels in the x-y plane and the z direction were clearly distinguished.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.