Abstract
The determination of trace levels of pesticides like atrazine in water samples of small, restricted volumes is one of the future demands of environmental analysis. In a brief review existing chromatographic and immunochemical methods for atrazine are critically discussed. Then a simple rapid enzyme-linked immunosorbent assay (ELISA) using the tip of an inoculation needle as a solid surface is presented. The sample volume could be reduced to 30 μl. The assay had a centre of the test IC 50 of 0.12 μg l −1 and permitted the characterisation of atrazine at levels of 0.022–2.90 μg l −1. A first outlook for automatisation is given. The new method was compared with an ELISA using 96 well microtiter plates as a solid phase. Surface water samples with low atrazine contents were analysed to check the new method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.