Abstract

We present the design of a variable-temperature ultrahigh vacuum (UHV) scanning tunneling microscope which can be operated between 20 and 400 K. The microscope is mounted directly onto the heat exchanger of a He continuous flow cryostat without vibration isolation inside the UHV chamber. The coarse approach is performed with an inertial slider driven by the same piezo tube that is also used for scanning. The performance of the instrument is demonstrated by two different kinds of measurements: First we show topographic images of close packed metal surfaces with atomic resolution. Then, we present results from scanning tunneling spectroscopy measurements on WSe2 under illumination at variable temperatures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.