Abstract

Alanine/EPR rods (3 × 10 mm) for routine use in high-dose radiation applications have been prepared by a simple technique in the laboratory where alanine powder was mixed with molten mixture of paraffin wax and ethylene vinyl acetate copolymer (EVA). The binding mixture EVA/Paraffin does not present interference or noise in the EPR signal before or after the irradiation. The rods show good mechanical properties for safe and multi-use handling. The rods can be used with good precision in the dose range from 1 to 125 kGy. The overall uncertainty for calibration of the EVAPA rod dosimeters at 2σ was found to be 4.56%. The dose response, influence of humidity and temperature during irradiation, energy dependence as well as post–irradiation storage at different conditions are discussed.