Abstract
Neutron streaming along the labyrinth of a generic bunker of an 18MV medical accelerator was evaluated. Monte Carlo simulations using MCNP code were performed to calculate neutron ambient dose equivalent along the labyrinth. The effect of plain, borated and barites concrete wall material, as well as borated concrete and plywood (Celotex), as neutron absorbing wall liners, was examined. The results of the study suggest that plywood can provide a cost effective material to attenuate neutron streaming along the labyrinth.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.