Abstract
An experimental study was carried out on the neutron response functions of two neutron detector arrays consisting of 39 3He proportional counters with a polyethylene moderator for monoenergetic neutrons within the 0.39–1.54 MeV neutron energy range. Experimental data on the sensitivity of neutron counting to a change in neutron energy and the influence of the thickness of polyethylene moderator were obtained. The experimental efficiency curves were compared with the calculated response functions generated by a neutron transport code.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.