Abstract

Measurement results are presented for the assay of plutonium in 220 l waste drums containing Pu-masses in the range 0.1–1g 240Pu eff obtained with the Time Interval Analysis (TIA) method. TIA is a neutron multiplicity method based on the concept of one- and two-dimensional Rossi-alpha distributions. The main source of measurement bias in neutron multiplicity measurements at low count-rates is the impredictable variation of the high-multiplicity neutron background of spallation neutrons induced by cosmic rays. The TIA-method was therefore equipped with a special background filter, which is designed and optimized to reduce the influence of these spallation neutrons by rejecting the high-multiplicity events. The measurement results, obtained with the background correction filter outlined in this paper, prove the repeatability and validity of the TIA-method and show that multiplicity counting with the TIA-technique is applicable for masses as low as 0.1 g 240Pu eff even at a detection efficiency of 12%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.