Abstract
A neutron multiplicity analysis system has been developed to determine the neutron multiplication of nonmoderating plutonium samples measured with thermal neutron counters and shift-register coincidence electronics. The system also determines the (α, n) neutron source strength and the effective 240Pu mass of the samples. The complete system is described and the measurement results are presented. Typical mass errors are ∼ 2% (1 σ). The principal limitation of the system is the long measurement time required for a sample with high plutonium mass or large (α, n) component.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods In Physics Research
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.