Abstract

The McCall method has been used to measure neutron leakage from the Mevatron 77, 18- and 15-MV photon beams. Gold foil activation has been used employing a beta counting technique for the 18-MV beam and a gamma counting technique for both the 18- and 15-MV beam. The two counting techniques were used to evaluate their relative merit. The measurements were made at various locations in the patient-treatment plane for different field sizes. The results show that the thermal-neutron dose equivalent contributes only about 1%-2% of the total neutron dose equivalent. At 100 cm, the neutron dose equivalent for the 18-MV beam is approximately six times that of the 15-MV beam, slightly exceeding the 0.1% of the useful beam criteria used by some of the regulatory agencies. In light of the uncertainty in fluence to dose equivalent conversion factors, the increased dose equivalent above 0.1% is insignificant.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.