Abstract

This study aims to measure the two-dimensional (2D) neutron spatial distribution inside a cubic and a head PMMA phantom for the purpose of further comparison with the treatment planning. The measurements were made by using the indirect neutron radiography (INR), which utilized a thin copper foil and the imaging plate. The developed image provides satisfactory spatial resolution and very low statistical error (< 1%). As to the time cost, the whole procedure normally takes less than 3 h. The result shows that the indirect neutron radiography can be a quick and reliable method to provide a 2D neutron spatial distribution inside a phantom.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call