Abstract

AbstractSubretinal injection is a complicated task for retinal surgeons to operate manually. In this paper we demonstrate a robust framework for needle detection and localisation in robot‐assisted subretinal injection using microscope‐integrated Optical Coherence Tomography with deep learning. Five convolutional neural networks with different architectures were evaluated. The main differences between the architectures are the amount of information they receive at the input layer. When evaluated on ex‐vivo pig eyes, the top performing network successfully detected all needles in the dataset and localised them with an Intersection over Union value of 0.55. The algorithm was evaluated by comparing the depth of the top and bottom edge of the predicted bounding box to the ground truth. This analysis showed that the top edge can be used to predict the depth of the needle with a maximum error of 8.5 μm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.