Abstract

Prompted by the need to support vibration and pressure measurements at frequencies down to 0.5 Hz (with expected future needs to 0.1 Hz), NBS now offers a calibration service for voltage standards and rms voltmeters in the range of 0.1-10 Hz. The means for the service is an Voltmeter/Calibrator, an NBS-developed instrument containing an rms digital voltmeter and ac and dc voltage calibrators. The methods used to calibrate the ac voltage calibrator are discussed; also, application of the ac Voltmeter/Calibrator to the calibration of customers' voltage and voltmeter standards is described. Finally, a multifrequency voltage reference source with frequency-independent amplitude is proposed as a more suitable transfer standard than thermal voltage converters (TVC's) for the 0.1-10-Hz range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.