Abstract
Hot tips are used either for characterizing nanostructures by using scanning thermal microscopes or for local heating to assist data writing. The tip-sample thermal interaction involves conduction at solid–solid contact as well as conduction through the ambient gas and through the water meniscus. We analyze those three heat transfer modes with experimental data and modeling. We conclude that the three modes contribute in a similar manner to the thermal contact conductance but they have distinct contact radii ranging from 30 nm to 1 μm. We also show that any scanning thermal microscope has a 1–3 μm resolution when used in ambient air.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.