Abstract
This work investigates the heat transfer between scanning thermal microscopy (SThM) probes and samples. It presents a detailed study of the heat transfer mechanisms that operate between the probe and the sample. Two SThM resistive probes of different sizes were used in active mode. Depending on the experimental conditions, the heat transferred to the sample through water meniscus, solid-solid contact and air is quantified. The methodology established to estimate the heat conduction through water meniscus shows that this mechanism is not dominant in the tip-sample heat exchange. Based on experimental results of measurements performed under vacuum conditions, the thermal boundary resistance at different contacts is estimated and in accordance with literature values. Through measurements performed under ambient conditions, the heat conduction through air appears to be strongly dependent on the sample thermal conductivity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.