Abstract

There are several kinds of medical laser treatment system, that is Nd:YAG, CO2, Argon Ion and etc., which are begun to use for animal experiments and clinical case.The important things to use appropriately the medical laser system are as follows.One is safety of laser system itself, another is method of maintenance, that is, safety control by user.In this paper, safety of laser system (YAG laser and CO2 laser) and how to control it were studied.As for hard-ware, it is concluded that following 4 items are thought to be necessary for safety of medical laser system1. System trouble is detectable before it is use.2. In case of appearance unusual conditions in the system, laser irradiation must be stopped automatically.(Interlock system)3. Alarm should be continued to ring during laser irradiation.4. In case of stand-by mode, whatever laser irradiation's switch become “ON”, laser beam must not be irradiated.In addition to these, general notices of use and safety control or soft-wareWare also reported in this paper.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.