Abstract

The characteristics of a newly developed axial-symmetric transmission gauge (AT gauge) for the measurement of wide pressure range were examined. The secondary electron multiplier used in the previous gauge was replaced by a Faraday cup for analog measurement of ion current. The lower limit of the pressure measurement at present was estimated to be about 10-11 Pa. The sensitivity factor of the AT gauge in the pressure range from 10-10 to 10-6 Pa calibrated by conductance modulation method and in the pressure range from 10-8 to 10-2 Pa calibrated by direct comparison with reference gauges were (2.3±0.04) ×10-3 Pa-1 for hydrogen and (5.8±0.1) ×10-3 Pa-1 for argon, respectively. Uncertainty of the ion current of the AT gauge by the fluctuation of the voltage 1 V of the energy filter was estimated to be about 1%. Long term stability of the electron emission current, intensity of gas phase ion and background signal was estimated to be about 0.03, 0.2 and 6.5%, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.