Abstract

In the SPring-8 Storage Ring, the pressure measured by Bayard-Alpert gauge near photon-absorber were not observed aging effects by a beam self-cleaning. As the result of investigations of this phenomenon, three causes of misreading of pressure measurement were found; (1) irradiation of the scattered X-ray to gauge head, (2) a large amount of electronic inflow to gauge head, (3) irradiation of the scattered X-ray to gauge cable. We took measures to the misreading, for example shielding of gauge heads and cables etc, and came to be able to measure pressure within tens of percent errors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.