Abstract

AbstractA magnetic separator attached in a zig zag flow channel assembly has been fabricated to detect myoglobin on the micro particles. An anti‐myoglobin conjugated magnetic hybrid and myoglobin conjugated silica micro particles were prepared and used for myoglobin detection on channel. For the detections, the competitive assay principle was followed based on affinity interaction of anti‐myoglobin‐myoglobin. The caliberation curve was plotted based on the deposit percentage of myoglobin conjugated silica micro particles. The calibration curve of the myoglobin showed the linear range between 1 × 10−8 ‐ 5 × 10−11 (M) (R2 = 0.9944). The minimum detectable concentration was 300 pg/mL. Hence all that concludes that, the application of zig zag flow channel with magnetic separator has offered new, simple and rapid approach for detecting myoglobin in whole blood samples with in 30 min.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.