Abstract

Multivariate standardization techniques [slope/bias (S/B) correction, single wavelength standardization (SWS) and piecewise direct standardization (PDS)] were used to attempt to correct changes over time in multivariate calibration models for potassium and calcium. These models were constructed with ion-selective electrode (ISE) arrays. Multivariate PDS local models which included the correlation between the sensors of the array were better than the other simple techniques. We considered the relationship between the variables (sensors) and, in the PDS treatment, we have indicated their arrangement which is taken from the loadings plot. We used the Kennard–Stone algorithm to select the standardization samples from the original responses of the samples and the partial least squares (PLS) scores of each model. These scores include information about the concentrations. The models and standardizations were validated by predictions on real samples such as natural waters. The best standardization conditions provided unbiased predictions with no loss of precision.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.