Abstract

An assisted liquid–liquid extraction of copper, iron, nickel and zinc from vegetable oil samples with subsequent determination by high-resolution continuum source flame atomic absorption spectrometry (HR-CS FAAS) was optimized by applying a full factorial design in two levels and the response surface methodology, Box–Behnken. The effects of the acid concentration and the amplitude, cycle and time of sonication on the extraction of the analytes, as well as their interactions, were assessed. In the selected condition (sonication amplitude=66%, sonication time=79s, sonication cycle=74%), using 0.5molL−1 HCl as the extractant, the limits of quantification were 0.14, 0.20, 0.21 and 0.04μgg−1 for Cu, Fe, Ni and Zn, respectively, with R.S.D. ranging from 1.4% to 3.6%. The proposed method was applied for the determination of the analytes in soybean, canola and sunflower oils.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call