Abstract

The aim of this work was to evaluate the microemulsification as sample preparation procedure for determination of Cu, Fe, Ni and Zn in vegetable oils samples by High-Resolution Continuum Source Flame Atomic Absorption Spectrometry (HR-CS FAAS). Microemulsions were prepared by mixing samples with propan-1-ol and aqueous acid solution, which allowed the use of inorganic aqueous standards for the calibration. To a sample mass of 0.5g, 100μL of hydrochloric acid and propan-1-ol were added and the resulting mixture diluted to a final volume of 10mL. The sample was manually shaken resulting in a visually homogeneous system. The main lines were selected for all studied metals and the detection limits (3σ, n=10) were 0.12, 0.62, 0.58 and 0.12mgkg−1 for Cu, Fe, Ni and Zn, respectively. The relative standard deviation (RSD) ranged from 5% to 11 % in samples spiked with 0.25 and 1.5μgmL−1 of each metal, respectively. Recoveries varied from 89% to 102%. The proposed method was applied to the determination of Cu, Fe, Ni and Zn in soybean, olive and sunflower oils.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call