Abstract

A multicolor quantum dot (QD)-based nanosensor for multiplex detection of two tumor markers in a homogeneous format based on fluorescence polarization immunoassay was proposed. QDs520 and QDs620 were labeled alpha-fetoprotein(α-AFP) and carcinoembryonic antigen (CEA), respectively. After separated and purified by ultrafiltration, they were used in fluorescence polarization immunoassay for the simultaneous detection of human serum alpha-fetoprotein and carcinoembryonic antigen. Under the optimal conditions, the multi-analyte immunosensor had a wide linear range (from 0.5ngmL−1 to 500ngmL−1) for both two tumor markers and good correlation (0.996 for α-AFP and 0.993 for CEA). The detection limits (LOD) were 0.36ngmL−1 for CEA and 0.28ngmL−1 for α-AFP (S/N=3). The carcinoembryonic antigen and fetoprotein in clinical serum samples were simultaneously detected. The results from 28 serum samples had a good agreement with enzyme-linked immunosorbent assay (ELISA). The relative standard deviation and the recovery suggested that the precision and the accuracy of this analytical method were satisfactory. This strategy with high sensitivity, good specificity, easy procedures and short analysis time shows great promise for clinical diagnoses and basic discovery. The application of QDs with longer fluorescence lifetime and small fluorescence polarization can be used for the determination of high molecular-weight substances which cannot be analyzed using dye fluorescence polarization immunoassay.