Abstract

BackgroundSaliva has garnered great interest as an alternative specimen type for molecular detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). Data are limited on the relative performance of different molecular methods using saliva specimens and the relative sensitivity of saliva to nasopharyngeal (NP) swabs.MethodsTo address the gap in knowledge, we enrolled symptomatic healthcare personnel (n = 250) from Barnes-Jewish Hospital/Washington University Medical Center and patients presenting to the Emergency Department with clinical symptoms compatible with coronavirus disease 2019 (COVID-19; n = 292). We collected paired saliva specimens and NP swabs. The Lyra SARS-CoV-2 assay (Quidel) was evaluated on paired saliva and NP samples. Subsequently we compared the Simplexa COVID-19 Direct Kit (Diasorin) and a modified SalivaDirect (Yale) assay on a subset of positive and negative saliva specimens.ResultsThe positive percent agreement (PPA) between saliva and NP samples using the Lyra SARS-CoV-2 assay was 63.2%. Saliva samples had higher SARS-CoV-2 cycle threshold values compared to NP swabs (P < 0.0001). We found a 76.47% (26/34) PPA for Simplexa COVID-19 Direct Kit on saliva and a 67.6% (23/34) PPA for SalivaDirect compared to NP swab results.ConclusionThese data demonstrate molecular assays have variability in performance for detection of SARS-CoV-2 in saliva.