Abstract

This paper describes a fully automated, multidimensional, double-oven GC-GC system, developed in our laboratory. The system is based on the use of mechanical valves which allow the multitransfer of different fractions during the same GC analysis and the use of the two GC independently when the multitransfer options is not used. Pneumatic and electronic circuits permit maintenance of constant retention times in the pre-column, even for the components eluted after numerous transfers. The system has been used for the determination of the enantiomeric distribution of β-pinene, sabinene, limonene, linalol, terpinen-4-ol, and α-terpineol in lemon oils. The results obtained allowed the characterization of cold-pressed lemon oils.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.