Abstract

The objectives of this study were to develop a multi-channel trans-impedance leadforming method for beat-to-beat stroke volume (SV) and breath-by-breath tidal volume (TV) measurements and assess its feasibility on an existing in vivo animal dataset. A deterministic leadforming algorithm was developed to extract a cardiac volume signal (CVS) and a respiratory volume signal (RVS) from 208-channel trans-impedance data acquired every 20 ms by an electrical impedance tomography (EIT) device. SVEIT and TVEIT values were computed as a valley-to-peak value in the CVS and RVS, respectively. The method was applied to the existing dataset from five mechanically-ventilated pigs undergoing ten mini-fluid challenges. An invasive hemodynamic monitor was used in the arterial pressure-based cardiac output (APCO) mode to simultaneously measure SVAPCO values while a mechanical ventilator provided TVVent values. The leadforming method could reliably extract the CVS and RVS from the 208-channel trans-impedance data measured with the EIT device, from which SVEIT and TVEIT were computed. The SVEIT and TVEIT values were comparable to those from the invasive hemodynamic monitor and mechanical ventilator. Using the data from 5 pigs and a simple calibration method to remove bias, the error in SVEIT and TVEIT was 9.5% and 5.4%, respectively. We developed a new leadforming method for the EIT device to robustly extract both SV and TV values in a deterministic fashion. Future animal and clinical studies are needed to validate this leadforming method in various subject populations. The leadforming method could be an integral component for a new cardiopulmonary monitor in the future to simultaneously measure SV and TV noninvasively, which would be beneficial to patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call