Abstract
AbstractThe peak parking (PP) method probes the longitudinal diffusion coefficient of a compound at a single location along the chromatographic column. We extended to a so-called multi-location peak parking (MLPP) method, in which a large number of axial locations along the column are selected in order to check the validity of the conventional PP method and to reveal possible defaults in the structure of the packed bed or pitfalls of the PP and the MLPP methods. MLPP was applied to a series of HILIC columns, including a 5.0μm Venusil, a 3.0μm Luna-diol, three 2.7μm Halo, and a 1.7μm Kinetex columns. The results demonstrate that the MLPP method may reveal local heterogeneities in the axial diffusion of small retained low molecular weight compounds along the column. Most importantly, experiments show that the sample zone should not be parked in the entrance of the column (i.e., at <1/10 th of the column length). The abrupt drop in the flow rate considerably affects the peak shape and prevents scientists from using the conventional PP method. Practical solutions to cope with that problem are proposed and their success/failure are discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.