Abstract

The combination of series of measurements of band broadening made with the peak parking (PP) method, using successively an open capillary tube and a HPLC column, gives a convenient procedure for the measurement of the molecular diffusivity ( D m) of compounds in solutions, of their axial dispersion coefficient ( D ax,m) in chromatographic columns, and of the tortuosity or obstructive factor of the column bed. The molecular diffusivity measured for benzene in methanol was in excellent agreement with literature data. The ratio of the axial dispersion coefficient to this diffusivity gives the obstructive factor ( γ m) of the packed bed, which was 0.74 for the column used. The values of D m in other solutions were obtained from the D ax,m values measured by the PP method, by correcting the D ax,m values with the γ m value. The D m values determined by this method were in good agreement with those previously reported or estimated using literature correlations. These results showed that the PP method is effective for the experimental measurement of D m.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.