Abstract

Very recently we have shown that CUBE-preamplifiers developed by XGLab s.r.l. can be used for the readout of single elements of thick structured planar HPGe- and Si(Li)-detectors produced by SEMIKON [1]. In this paper we will present the results of a simultaneous multi-element readout of structured detectors using the same preamplifiers for measuring high-energy x-rays (more than 100 keV) with a comparable energy resolution as for the single-element readout. Several high-purity germanium detectors (HPGe-detectors) with different position sensitive structures on one detector contact have been used for the first tests. In addition to that we have modified an existing 16-pixel HPGe-polarimeter from GSI-Darmstadt with the new readout. The detector elements (7 mm × 7 mm each, arranged in a 4 × 4 matrix) are connected to CUBE-preamplifiers used in pulse-reset mode. The technological progress achieved with this detector system resulting in a significant improved energy resolution will contribute a lot to much more precise polarization measurements of x-rays emitted from atom-ion collisions which are part of the physics program of the SPARC collaboration (Stored Particles Atomic Physics Research Collaboration) at GSI and the future FAIR accelerator facility (Facility for Antiproton and Ion Research).