Abstract

Details of the design and construction of a two-channel air-fuel ratio (AFR) chromatograph for the on-line analysis of a single sample of wet combustion gases are given. A thermal conductivity detector (TCD) is used to detect CO 2 and N 2 components, which are separated on a single column system. Then the ratio of the concentration of CO 2 and N 2 is used to determine the AFR in the original sample. The column used is a stainless-steel tube ( 3 in. × 1 8 in. I.D. ) filled with 80 mesh Porapak Q. A switching valve is incorporated to divert the samples between the separating column. The instrument is fully automatic and capable of determining two different AFR measurements in a working time of less than 5 s. It can operate satisfactorily at mixture strengths that are well above stoichiometric levels.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.