Abstract

Type E thermocouples show magnetic effects at liquid nitrogen temperature and below. This may cause trouble in experiments that are sensitive to magnetic stray fields like nuclear magnetic resonance, photoemission or high resolution electron energy loss spectroscopy. Here, a solution for the temperature measurement of a single crystal is presented. The authors weld a copper rod onto the back side of the single crystal, thereby relocating the sensitive sample from the thermocouple attachment position. They show that it is possible to measure the crystal temperature at the end of the rod while significantly reducing the ferromagnetic influence due to the increased distance.