Abstract

The LB 1000-PW detector is mainly used for determining total alpha, beta and gamma activity of low activity natural sources such as water, soil, air filters and any other environmental sources. Detector efficiency needs to be known in order to measure the absolute activity of such samples. This paper presents results on the total gamma counting efficiency of a Phoswich detector from point and disk sources. The directions of photons emitted from the source were determined by Monte Carlo techniques and the true path lengths in the detector were determined by analytical equations depending on photon directions. Results are tabulated for various gamma energies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.