Abstract
The Vero4DRT (Brainlab AG) linear accelerator is capable of dynamic tumor tracking (DTT) by panning/tilting the radiation beam to follow respiratory-induced tumor motion in real time. In this study, the panning/tilting motion is modeled in Monte Carlo (MC) for quality assurance (QA) of four-dimensional (4D) dose distributions created within the treatment planning system (TPS). Step-and-shoot intensity-modulated radiation therapy plans were optimized for 10 previously treated liver patients. These plans were recalculated on multiple phases of a 4D computed tomography (4DCT) scan using MC while modeling panning/tilting. The dose distributions on each phase were accumulated to create a respiratory-weighted 4D dose distribution. Differences between the TPS and MC modeled doses were examined. On average, 4D dose calculations in MC showed the maximum dose of an organ at risk (OAR) to be 10% greater than the TPS' three-dimensional dose calculation (collapsed cone [CC] convolution algorithm) predicted. MC's 4D dose calculations showed that 6 out of 24 OARs could exceed their specified dose limits, and calculated their maximum dose to be 4% higher on average (up to 13%) than the TPS' 4D dose calculations. Dose differences between MC and the TPS were greatest in the beam penumbra region. Modeling panning/tilting for DTT has been successfully modeled with MC and is a useful tool to QA respiratory-correlated 4D dose distributions. The dose differences between the TPS and MC calculations highlight the importance of using 4D MC to confirm the safety of OAR doses before DTT treatments.