Abstract

The theoretical properties of a monoexponential flow index, analogous to the one used earlier by other investigators for regional CBF (rCBF) measured after intraarterial injection, were investigated after the administration of 133Xe intraarterially, intravenously, and by inhalation under high and low flow conditions. The sensitivity of the flow index to changes in fast flow components or changes in the weight ratio between the fast and the slow flow compartments was found to be dependent on whichever part of the 133Xe clearance curve was used for the flow calculation and on the shape of the input function for 133Xe. Since biexponential analysis of the clearance curves includes a monoexponential approximation for each of the two components of the clearance curve corresponding to the high and the low flow "families" in the brain, the limitations of the monoexponential flow index observed are in principle also valid for the results of biexponential analysis of the clearance curves.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.