Abstract
In a prospective study we evaluated a novel fluorescence polarization immunoassay (FPIA) for determining cyclosporine (CsA) in whole blood. FPIA uses a monoclonal antibody and is performed on the TDx (Abbott). The within-series (CV less than 2%) and between-days (CV less than 3.3%) precision of the assay was excellent. The results obtained by the monoclonal FPIA in samples from transplant patients (n = 100) averaged 31.9% and 20.2% higher than those by HPLC and a specific radioimmunoassay (INCStar), respectively. Results by all three methods correlated well. Follow-up studies during the early course after liver transplantation, however, suggested that high metabolite concentrations affect FPIA results. This is explained by previously described cross-reactions of the monoclonal antibody with some CsA metabolites. The FPIA results in samples of such patients should be interpreted cautiously.