Abstract

This study compared the analysis of whole blood cyclosporine concentrations measured by fluorescence polarization immunoassay (FPIA) and radioimmunoassay (RIA) polyclonal and monoclonal procedures. Fifteen orthotopic liver transplant patients with a mean age of 39 +/- 11.06 years were included in the study. One hundred thirteen levels were analyzed using FPIA, RIA polyclonal, and RIA monoclonal procedures. There was no difference statistically in comparing FPIA and RIA polyclonal results (p greater than 0.05). There was a statistical difference between FPIA and RIA monoclonal results p = 0.0001). With use of least squares simple linear regression analysis, FPIA results showed good correlation with RIA polyclonal results (R2 = 0.87). Poor correlation was shown between FPIA and RIA monoclonal results (R2 = 0.51). In this study population, FPIA produced results 2.5% higher than the RIA polyclonal procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.