Abstract

PurposeTo present details about the innovative and disruptive technology of monochromatic X-rays and its application to breast imaging. MethodsTo analyze results of studies done using a prototype system for breast imaging that generates monochromatic X-rays through fluorescence emission. To assess signal-to-noise ratio (SNR) as a measure of image quality at different doses in breast phantoms of different sizes and review the comparison of parameters with a standard mammography system. ResultsMonochromatic X-rays reduce the radiation dose per mammogram by a factor of 5 to 10 times. For phantom simulating thick breast (9 cm), the SNR for monochromatic system was 2.6 times higher and the dose 4.2 times lower than the respective values obtained with the conventional system within the same 5 mm × 5 mm square area of the 100% glandular step wedge. For the conventional broadband system to equal the SNR of the monochromatic system, it would require a dose of 19 mGy, 29 times higher than the dose delivered by the monochromatic system. Contrast-enhanced digital mammography with monochromatic X-rays is shown to provide a simpler and more effective technique at substantially lower radiation dose. ConclusionsLowering radiation dose by a factor of 5 to 10 while maintaining image quality implies a major reduction in total exposure from breast cancer screening and dramatically less risk of radiation-induced cancers in at-risk women. The high SNRs for very thick breast phantoms provide strong evidence that screening with lower breast compression is possible while maintaining image quality.