Abstract

75 μm thick PTFE tape attracts interest, forming the basis of a potential skin dosimeter capable of utilization in radiotherapeutic external beam applications. The medium is readily available, for instance in the form of ordinary plumber's tape as used herein, typically of 12 mm width mounted on a spool. A60Co (mean energy 1.25 MeV) irradiator has been used to deliver doses in the range 0.1- to 10 Gy, with disrupted molecular bonds inspected via Raman spectrometry. The peak of greatest magnitude, at a wavenumber of ∼734 cm−1, corresponds to the difluoromethylene (CF2) symmetric stretching mode. Reduction in the peak intensity is seen to occur over the range 0.1- to 5 Gy, covering the values of skin dose typically arising in a single fractionated radiotherapeutic treatment. Results from these preliminary investigations point to a technique offering viability for radiotherapy skin dosimetry, utilizing a soft, low atomic number medium of particularly low-cost, also available in hypoallergenic form.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.