Abstract

The modulation transfer function (MTF) is a fundamental tool for assessing the sharpness of digital breast tomosynthesis (DBT) systems and is primarily measured using edge devices. We compared the MTF of a Senographe Pristina DBT system using four-edge devices. These devices were composed of stainless steel with a thickness of 0.6, 0.8, and 1.0mm, and 1.0mm tungsten, based on different international guidelines. We evaluated spatial frequencies at MTFs of 0.5 (MTF50%) and 0.1 (MTF10%). The collimator-equipped and non-collimator configurations of the DBT were compared. We found no appreciable differences between scan and chest wall-nipple directions. Both MTF50% (2.90-2.99 cycles/mm) and MTF10% (6.69-6.94 cycles/mm) demonstrated minimal variation across the different edge devices. The collimator-equipped system exhibited an MTF50% that was approximately 5% higher than that of the non-collimator configuration. The choice of the edge device did not appreciably impact the MTF.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.