Abstract

This note describes the design, construction, and test results of a modified sample holder used during low-temperature deep-level transient spectroscopy, current-voltage and capacitance-voltage measurements. This improved sample holder allows temperature scan rates of up to 6 K/min with a temperature shift of less than 1 K. High electrical isolation makes this sample holder also suitable for low-temperature current-voltage and capacitance-voltage measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.