Abstract

Flow injection amperometric quantification of ascorbic acid (AA), dopamine (DA), epinephrine (EP) and dipyrone (DI) in mixtures (in the microgram g-1 range) was successfully performed by using an array of microelectrodes with units modified by the electrodeposition of different noble metals, together with multivariate calibration analysis. The four groups of microelectrodes utilized included a pure gold electrode and electrodes modified by electrodeposition of platinum, palladium or a mixture of platinum + palladium. The array of microelectrodes was inserted in a flow cell and the amperometric data acquisition was performed with a four-channel potentiostat. The analysis of the resulting signals was carried out by a multivariate calibration method, using a group of 16 standard mixtures selected by a two-level factorial design. The analysis of synthetic samples and pharmaceutical compounds containing AA and DI led to very similar values to those obtained by the classical iodimetric analysis. The average absolute errors (in microgram g-1) calculated for each analyte were 0.3, 0.2, 0.4 and 0.4 for AA, DA, EP and DI, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.