Abstract
Bilirubin interferes with the Du Pont magnesium method in the aca at 510 nm. We determined that bilirubin concentrations in serum samples up to 380 mg/L did not affect the absorbance measured in the aca III at 540 nm, and therefore we modified the Du Pont setting of spectrophotometer for the magnesium method to 540 and 600 nm. Accuracy and precision of the modified method were comparable with the unmodified method and with atomic absorption spectrophotometry. For comparison, we analyzed serum samples with normal (n = 37) and increased (n = 22) bilirubin concentration with the modified method in the aca III (y) and in the atomic absorption spectrophotometer (x). The results (range 0.56-1.25 mmol/L) were in good agreement (x = y = 0.84 mmol/L, D(x - y) = -0.001, SD = 0.004, t = 0.238, P greater than 0.50) and bilirubin did not interfere with the modified method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.