Abstract
The End-Systolic Pressure-Volume Relation (ESPVR) is generally modelled as a linear relationship between P and V as cardiac reflexes, such as the baroreflex, are typically suppressed in experiments. However, ESPVR has been observed to behave in a curvilinear fashion when cardiac reflexes are not suppressed, suggesting the curvilinear function may be more clinically appropriate. Data was gathered from 41 vena cava occlusion manoeuvres performed experimentally at a variety of PEEPs across 6 porcine specimens, and ESPVR determined for each pig. An exponential model of ESPVR was found to provide a higher correlation coefficient than a linear model in 6 out of 7 cases, and a lower Akaike Information Criterion (AIC) value in all cases. Further, the exponential ESPVR provided positive V0 values in a physiological range in 6 out of 7 cases analysed, while the linear ESPVR produced positive V0 values in only 3 out of 7 cases, suggesting linear extrapolation of ESPVR to determine V0 may be flawed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.