Abstract

Beyond the conventional non-contact temperature measurement with ambient temperature compensation and our previous work of multi-sensors compensation, two compensation schemes are proposed and compared. An infrared radiation measurement module with several temperature sensors are built and calibrated with ambient compensation to investigate the dynamic temperature distribution under moving of module from one place to another. The new approach we proposed including the infrared radiation exchanges model between the target, sensor and the optical path thorough temperature monitoring and two fast temperature measurement schemes with dynamic compensation. After careful calibrations and verification of several experiment conditions, our models of two dynamic compensation schemes both show an excellent agreement with the measuring data. The experimental data of compensation scheme reach a stable reading value of target temperature down from 40 min to 2 min for the differential scheme compensation and 6 min for multi-sensors scheme compensation with temperature error around 0.2 °C.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.