Abstract

The new trend towards minimally invasive mm-sized and free-floating distributed implants imposes a significant challenge in wireless power transmission to these medical devices. The magnetic field produced by external transmitter (Tx) coils at the position of the implants can be considered homogeneous to separate optimization of the Tx and receiver (Rx) coils for efficient power transfer. This paper focuses on optimization of the solenoid-type Rx coils, which are suitable for this application. We have developed an analytical model of solenoid coils that includes the impact of tissue and coating around the coils and verifies through simulations and measurements. Using the proposed model, under a given size restriction and a specific load, we find the optimal operating frequency and coil geometry by maximizing a figure of merit for Rx. For a mm-sized coil, the optimal operating frequency and the number of turns are found 500 MHz and 6, respectively, if the coil is closely wound using AWG36 copper wires.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.